Diabetes Care, Vol 20, Issue 7 1114-1121, Copyright © 1997 by American Diabetes Association
Open-flow microperfusion of subcutaneous adipose tissue for on-line continuous ex vivo measurement of glucose concentration
Z Trajanoski, GA Brunner, L Schaupp, M Ellmerer, P Wach, TR Pieber, P Kotanko and F Skrabal
Department of Biophysics, Institute of Biomedical Engineering, Graz University of Technology, Austria. trajanoski@ibmt.tu-graz.ac.at
OBJECTIVE: To evaluate a novel technique for on-line continuous glucose
measurement in subcutaneous adipose tissue, and to investigate its accuracy
for detection of hypoglycemia. RESEARCH DESIGN AND METHODS: The method
combined an open-flow microperfusion of subcutaneous adipose tissue using a
double lumen catheter and an extracorporeal sensor cell. An isotonic
ion-free solution was perfused through the inner lumen of the catheter,
equilibrated with the subcutaneous tissue fluid, and sampled through the
outer lumen. The recovery was continuously monitored as the ratio between
the measured sampled fluid conductivity and the subcutaneous tissue fluid
conductivity (assumed to have a constant value of 1.28 S/m at 25 degrees
C). Glucose concentration was calculated on-line from the measured glucose
in the sampled fluid and the measured recovery in healthy volunteers during
hyperglycemic glucose loads (n = 8), hypoglycemic hyperinsulinemic clamp (n
= 6), and a 24-h monitoring period (n = 7). RESULTS: Subcutaneous glucose
concentrations in the fasting state were 94% of the plasma glucose
concentrations in arterialized venous samples. According to the error grid
analysis, 96.9% of the on-line measured subcutaneous glucose concentrations
during hyperglycemia and 96.3% during hypoglycemia were in accurate or
acceptable zones. The mean differences between the measured subcutaneous
glucose and the actual plasma glucose concentration were -0.06-3.3 mmol/l
(hyperglycemia), and -0.6-1.1 mmol/l (hypoglycemia). CONCLUSIONS: By
combining open-flow microperfusion, glucose sensor, and conductivity
measurement, glucose concentration in the subcutaneous adipose tissue can
be monitored on-line, extracorporeally, and continuously without any in
vivo calibration, and gives accurate measurements during hyper- and
hypoglycemia.

CiteULike Del.icio.us Digg Reddit Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
C. Pachler, D. Ikeoka, J. Plank, H. Weinhandl, M. Suppan, J. K. Mader, M. Bodenlenz, W. Regittnig, H. Mangge, T. R. Pieber, et al.
Subcutaneous adipose tissue exerts proinflammatory cytokines after minimal trauma in humans
Am J Physiol Endocrinol Metab,
September 1, 2007;
293(3):
E690 - E696.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Weinhandl, C. Pachler, J. K. Mader, D. Ikeoka, A. Mautner, A. Falk, M. Suppan, T. R. Pieber, and M. Ellmerer
Physiological Hyperinsulinemia Has No Detectable Effect on Access of Macromolecules to Insulin-Sensitive Tissues in Healthy Humans
Diabetes,
September 1, 2007;
56(9):
2213 - 2217.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Ellmerer, M. Haluzik, J. Blaha, J. Kremen, S. Svacina, W. Toller, J. Mader, L. Schaupp, J. Plank, and T. Pieber
Clinical Evaluation of Alternative-Site Glucose Measurements in Patients After Major Cardiac Surgery.
Diabetes Care,
June 1, 2006;
29(6):
1275 - 1281.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
W. Regittnig, M. Ellmerer, G. Fauler, G. Sendlhofer, Z. Trajanoski, H.-J. Leis, L. Schaupp, P. Wach, and T. R. Pieber
Assessment of transcapillary glucose exchange in human skeletal muscle and adipose tissue
Am J Physiol Endocrinol Metab,
August 1, 2003;
285(2):
E241 - E251.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Maran, C. Crepaldi, A. Tiengo, G. Grassi, E. Vitali, G. Pagano, S. Bistoni, G. Calabrese, F. Santeusanio, F. Leonetti, et al.
Continuous Subcutaneous Glucose Monitoring in Diabetic Patients: A multicenter analysis
Diabetes Care,
February 1, 2002;
25(2):
347 - 352.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Ellmerer, L. Schaupp, G. A. Brunner, G. Sendlhofer, A. Wutte, P. Wach, and T. R. Pieber
Measurement of interstitial albumin in human skeletal muscle and adipose tissue by open-flow microperfusion
Am J Physiol Endocrinol Metab,
February 1, 2000;
278(2):
E352 - E356.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. W. Cline, K. F. Petersen, M. Krssak, J. Shen, R. S. Hundal, Z. Trajanoski, S. Inzucchi, A. Dresner, D. L. Rothman, and G. I. Shulman
Impaired Glucose Transport as a Cause of Decreased Insulin-Stimulated Muscle Glycogen Synthesis in Type 2 Diabetes
N. Engl. J. Med.,
July 22, 1999;
341(4):
240 - 246.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. Schaupp, M. Ellmerer, G. A. Brunner, A. Wutte, G. Sendlhofer, Z. Trajanoski, F. Skrabal, T. R. Pieber, and P. Wach
Direct access to interstitial fluid in adipose tissue in humans by use of open-flow microperfusion
Am J Physiol Endocrinol Metab,
February 1, 1999;
276(2):
E401 - E408.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Ellmerer, L. Schaupp, G. Sendlhofer, A. Wutte, G. A. Brunner, Z. Trajanoski, F. Skrabal, P. Wach, and T. R. Pieber
Lactate Metabolism of Subcutaneous Adipose Tissue Studied by Open Flow Microperfusion
J. Clin. Endocrinol. Metab.,
December 1, 1998;
83(12):
4394 - 4401.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
G. W. Cline, B. M. Jucker, Z. Trajanoski, A. J. M. Rennings, and G. I. Shulman
A novel 13C NMR method to assess intracellular glucose concentration in muscle, in vivo
Am J Physiol Endocrinol Metab,
February 1, 1998;
274(2):
E381 - E389.
[Abstract]
[Full Text]
[PDF]
|
 |
|
Copyright © 1997 by the American Diabetes Association.
|
|
| |
|