Diabetes Care
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Purchase Article
Right arrow View Shopping Cart
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lteif, A. N.
Right arrow Articles by Schwenk, W. F.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lteif, A. N.
Right arrow Articles by Schwenk, W. F.
Social Bookmarking
 Add to CiteULike   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

Diabetes Care, Vol 22, Issue 1 137-140, Copyright © 1999 by American Diabetes Association


ARTICLES

Accuracy of pen injectors versus insulin syringes in children with type 1 diabetes

AN Lteif and WF Schwenk
Department of Pediatric Endocrinology, Mayo Clinic, Rochester, Minnesota 55905, USA.

OBJECTIVE: To compare the accuracy and precision of insulin syringes and pen devices used by children with type 1 diabetes and their parents. RESEARCH DESIGN AND METHODS: There were 48 subjects (32 patients, a parent of an additional 16 patients) instructed to measure out morning insulin doses three times from vials and/or cartridges containing saline mixed with small amounts of [14C]glucose (solution used as regular insulin) and [3H]glucose (solution used as NPH insulin) and to dispense the contents into a scintillation vial. Statistical analysis was used to determine the accuracy and precision of both methods of insulin delivery. RESULTS: The absolute error in measuring out doses of regular insulin < 5 U was greater with insulin syringes compared with pen injection devices (9.9 +/- 2.4 vs. 4.9 +/- 1.6%, respectively). Both were comparable for regular insulin doses > 5 U (3.2 +/- 0.6 vs. 2.2 +/- 0.4% for syringes and pens, respectively). The accuracy in drawing up NPH doses was similar for low and high insulin doses (mean percent error of 7.5 +/- 1.5 vs. 5.6 +/- 1.1%). CONCLUSIONS: Pen devices are more accurate than insulin syringes in measuring out insulin at low insulin doses. The accuracy of insulin syringes improves when higher doses of regular insulin are measured out and becomes comparable to pen devices.