Diabetes Care 26:582-589, 2003
© 2003 by the American Diabetes Association, Inc.
Clinical Care/Education/Nutrition Original Article |
Clinical Performance of CGMS in Type 1 Diabetic Patients Treated by Continuous Subcutaneous Insulin Infusion Using Insulin Analogs
Bruno Guerci, MD, PHD,
Michèle Floriot, MD,
Philip Böhme, MD,
Danielle Durain, RN,
Muriel Benichou, MD,
Stéphanie Jellimann, MD and
Pierre Drouin, MD*,
Service de Diabétologie, Maladies Métaboliques et Maladies de la Nutrition, Centre dInvestigation Clinique-INSERM/CHU de NANCY-Hôpital Jeanne dArc, Centre Hospitalo-Universitaire de Nancy, Toul cedex, France
OBJECTIVEPerformance criteria have been established for in vitro blood glucose monitoring, particularly for the self-monitoring of blood glucose using glucose meters. Devices intended for use in the future, such as the continuous glucose monitoring system (CGMS), should satisfy similar criteria, particularly in diabetic patients under intensive therapy.
RESEARCH DESIGN AND METHODSThe analysis was conducted on 18 type 1 diabetic patients (not controlled, HbA1c >7.5%) treated by external pump using insulin analogs. Each patient received a glucose sensor for 3 days during his/her hospitalization and was instructed in its operation. Medtronic criteria were used to determine the accuracy of the CGMS. In addition, the data were analyzed according to American Diabetes Association (ADA) criteria, Clarke Error Grid analysis, and method of residuals, with the glucose oxidase method using a Beckman analyzer used as the reference method. Specificity and sensitivity were evaluated from the viewpoint of accuracy in the detection of hypoglycemia. For nine patients, two glucose sensors were simultaneously inserted into an abdominal site to determine the reproducibility of the system.
RESULTSAmong the 33 glucose sensors inserted, 6 (18%) were nonfunctional. The mean duration of CGMS recording was 63 ± 12 h. From all of the 692 sets of data that paired glucose readings and CGMS, the coefficients of correlation ranged from 0.87 to 0.92 and the mean absolute error ranged from 12.8 to 15.7%. The time experienced in hypoglycemia (<55 mg/dl) was reported at 86 ± 62 min/day. Only 39% of the CGMS values satisfied the ADA precision criteria to within ±10%, and 19% of these values satisfied the future ADA precision criteria of accuracy to within ±5%. The means of difference method showed that the CGMS slightly underestimated the plasma glucose values (mean = -12 mg/dl). Error grid analysis showed only 77% of the glucose sensor values were in zone A, and 98.9% were in zones A and B. Two values fell in zone C and a single value fell in zone D. The sensitivity and specificity of the CGMS to detect hypoglycemia were 33 and 96%, respectively. A total of 6,666 paired sensor values were recorded with a coefficient of correlation of 0.84 with a coefficient of variation of 8.25%.
CONCLUSIONSCGMS could be useful in routine clinical practice to provide much more information on the glucose profile than intermittent self-monitoring of blood glucose (SMBG). However, CGMS cannot be used as a replacement for glucose meters because it does not satisfy the conventional performance goals set down for in vitro glucose measurements and could therefore lead to clinically incorrect treatment decisions.
Abbreviations: ADA, American Diabetes Association CGMS, continuous glucose monitoring system FN, false negative FP, false positive SMBG, self-monitoring of blood glucose TN, true negative TP, true positive

CiteULike Del.icio.us Digg Reddit Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
M. R. Burge, S. Mitchell, A. Sawyer, and D. S. Schade
Continuous Glucose Monitoring: The Future of Diabetes Management
Diabetes Spectr,
April 1, 2008;
21(2):
112 - 119.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. G. Piper, J. L. Alexander, A. Shukla, F. Pigula, J. M. Costello, P. C. Laussen, T. Jaksic, and M. S.D. Agus
Real-Time Continuous Glucose Monitoring in Pediatric Patients During and After Cardiac Surgery
Pediatrics,
September 1, 2006;
118(3):
1176 - 1184.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Yates, A. Hasnat Milton, K. Dear, and G. Ambler
Continuous Glucose Monitoring-Guided Insulin Adjustment in Children and Adolescents on Near-Physiological Insulin Regimens: A randomized controlled trial
Diabetes Care,
July 1, 2006;
29(7):
1512 - 1517.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. Monnier, E. Mas, C. Ginet, F. Michel, L. Villon, J.-P. Cristol, and C. Colette
Activation of Oxidative Stress by Acute Glucose Fluctuations Compared With Sustained Chronic Hyperglycemia in Patients With Type 2 Diabetes
JAMA,
April 12, 2006;
295(14):
1681 - 1687.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
I. M. Wentholt, M. A. Vollebregt, A. A. Hart, J. B. Hoekstra, and J. H. DeVries
Comparison of a Needle-Type and a Microdialysis Continuous Glucose Monitor in Type 1 Diabetic Patients
Diabetes Care,
December 1, 2005;
28(12):
2871 - 2876.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. C. Klonoff
Continuous Glucose Monitoring: Roadmap for 21st century diabetes therapy
Diabetes Care,
May 1, 2005;
28(5):
1231 - 1239.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. S. Jeha, L. P. Karaviti, B. Anderson, E. O'Brian Smith, S. Donaldson, T. S. McGirk, and M. W. Haymond
Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes
Diabetes Care,
December 1, 2004;
27(12):
2881 - 2886.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Accuracy of the GlucoWatch G2 Biographer and the Continuous Glucose Monitoring System During Hypoglycemia: Experience of the Diabetes Research in Children Network
Diabetes Care,
March 1, 2004;
27(3):
722 - 726.
[Abstract]
[Full Text]
[PDF]
|
 |
|
Copyright © 2003 by the American Diabetes Association.
|
|
| |
|