Comparison of Insulin Aspart With Buffered Regular Insulin and Insulin Lispro in Continuous Subcutaneous Insulin InfusionA randomized study in type 1 diabetes
1 Atlanta Diabetes Associates, Atlanta, Georgia
OBJECTIVETo compare the safety and efficacy of insulin aspart (IAsp), buffered regular insulin (BR), and insulin lispro administered by continuous subcutaneous insulin infusion (CSII) in patients with type 1 diabetes. RESEARCH DESIGN AND METHODSAfter completing a 4-week run-in period with BR, 146 adult patients with type 1 diabetes (with pretrial CSII experience) were randomly assigned (2:2:1) to CSII treatment with IAsp, BR, or lispro for 16 weeks in a multicenter, open-label, randomized, parallel-group study. Bolus insulin doses were administered 30 min before meals (BR) or immediately before meals (IAsp or lispro).
RESULTSTreatment groups had similar baseline HbA1c (7.3% ±0.7 for IAsp, 7.5% ±0.8 for BR, and 7.3% ±0.7 for lispro). After 16 weeks of treatment, HbA1c values were relatively unchanged from baseline, and the mean changes in baseline HbA1c values were not significantly different between the three groups (0.00 ±0.51, 0.15 ±0.63, and 0.18 ±0.84 for the IAsp, BR, and lispro groups, respectively). The rates of hypoglycemic episodes (blood glucose <50 mg/dl) per patient per month were similar (3.7, 4.8, and 4.4 for the IAsp, BR, and lispro groups, respectively). Clogs/blockages in pumps or infusion sets were infrequent; most subjects (76, 83, and 75% in the IAsp, BR, and lispro groups, respectively) had CONCLUSIONSInsulin aspart in CSII was as efficacious and well tolerated as BR and lispro and is a suitable insulin for continuous subcutaneous insulin infusion using external pumps.
Abbreviations: BG, blood glucose BR, buffered regular insulin CSII, continuous subcutaneous insulin infusion DCCT, Diabetes Control and Complications Trial IAsp, insulin aspart MDI, multiple daily injection SD, standard deviation
Continuous subcutaneous insulin infusion (CSII) therapy offers the potential of controlled basal insulin release, which can closely mimic the insulin profile of individuals without diabetes. In the Diabetes Control and Complications Trial (DCCT), patients on intensive insulin therapy could choose between CSII therapy and multiple daily injection (MDI) therapy and could change between therapies during the trial (1). Although the DCCT was not specifically designed to compare CSII and MDI therapies, the mean HbA1c value of subjects choosing CSII therapy for 90% of the time was 0.2% less than the HbA1c value of the MDI therapy group (2). CSII therapy has also been shown to decrease the incidence of severe hypoglycemia as compared with MDI therapy; this benefit was attributed to the greater reproducibility and flexibility of insulin administration during CSII therapy (3). These advantages of CSII, as well as improvements in pump technology, have led to increasing acceptance of insulin pump therapy (4). Insulin aspart (IAsp) has properties that are expected to be optimal for use in CSII: soluble, rapid-acting, and uniform absorption characteristics (5). Such properties should reduce the size of the subcutaneous insulin depot and reduce the time interval between insulin administration and insulin action. Because the concept of CSII is predicated on an immediacy of insulin action, a rapid-acting insulin analog should be ideal. Studies using such an analog, lispro, have validated this concept (6,7). As currently formulated, IAsp is physically compatible for use in pumps and has been shown to be as safe and effective as buffered regular insulin (BR) in CSII therapy (8). The only insulin currently approved for CSII use in the U.S. is BR, although insulin lispro is commonly used in CSII. The present study was conducted to further examine the suitability of IAsp for CSII therapy by directly comparing its safety and efficacy in type 1 diabetes with that of BR and lispro.
This was an open-label, randomized, parallel-group study conducted at 13 sites in the U.S. Patients with type 1 diabetes received IAsp, BR, or lispro as a continuous subcutaneous infusion by an external pump for 16 weeks. The study was performed in accordance with the Declaration of Helsinki and with the approval of local independent review boards. Written informed consent was obtained from all subjects.
Subjects
Treatments After the run-in period, subjects were randomly assigned (2:2:1) to the lowest available randomization number, to receive either IAsp, BR, or lispro (Humalog; Eli Lilly, Indianapolis, IN). The randomization code was provided by Novo Nordisk A/S (Bagsvaerd, Denmark) to ensure that the investigator and subject were blinded at the point of randomization. Subjects assigned to the IAsp or lispro groups were instructed to take bolus doses just before the start of each meal, whereas those assigned to the BR treatment group were instructed to take bolus doses 30 min before the start of each meal. During the run-in period and during the first 4 weeks after randomization (dose-adjustment period), the investigator reviewed the blood glucose meter readings with the subject to maximize drug therapy to achieve the targeted fasting (prebreakfast) blood glucose level between 80 and 120 mg/dl without unacceptable hypoglycemia. Subjects were contacted at least weekly during the run-in period and every 2 weeks during the dose-adjustment period. Telephone contact with the clinic could be made by the subjects in the event of problems with the treatment or by the investigator if insulin doses were modified. The subjects continued on the adjusted dose regimen during weeks 416 (maintenance period) unless further dose adjustment was required.
Efficacy assessments
Safety assessments Subjects were made aware of the possibility of clogs or blockages of the pump or infusion set and were told to record these events in their diaries.
Statistical analysis
Subjects Baseline demographic characteristics were similar for all treatment groups (Table 1). Overall, 93, 85, and 96% of the subjects receiving IAsp, BR, and lispro, respectively, completed the 16-week study.
Subjects had a mean baseline insulin requirement of 0.7, 0.6, and 0.5 units/kg in the IAsp, BR, and lispro groups, respectively, which did not change by the end of the study (Table 1). In general, subjects maintained the same insulin dose (basal and bolus) throughout the study.
Efficacy Mean eight-point BG profiles at baseline and at end of study are presented in Fig. 1. Postprandial values for subjects in the rapid-acting insulin analog groups were improved from baseline values and tended to be lower than those for subjects in the BR group. A few statistically significant differences were observed at week 16 between the treatment groups: i.e., at dinner + 90 min, the BG value for the IAsp group was lower than those for BR and lispro groups (P=0.019); at 2:00 A.M., the BG value for the BR group was lower than those for IAsp and lispro groups (P=0.002).
The mean values for HDL, LDL, cholesterol, and triglycerides were within their normal limits at baseline and at week 16 for each treatment group. Mean changes in baseline values for lipid parameters were not significantly different between treatment groups. Similarly, mean weights at baseline were comparable for each group and did not significantly increase or decrease during the study.
Hypoglycemia
Of the confirmed episodes, the rate of nocturnal hypoglycemic episodes (midnight to 6:00 ) for the IAsp group was lower than that for the BR group and similar to that of the lispro group (Table 2). During the 3-month maintenance period, a greater percentage of patients in the IAsp group (41%, 24/59) were free of nocturnal hypoglycemic episodes compared with patients in the BR (20%, 12/59) or lispro (25%, 7/28) groups. No major nocturnal hypoglycemic episodes occurred during the study.
Safety Hyperglycemia (BG >350 mg/dl) was the most commonly reported adverse event for each treatment group and was reported by 27% (16/59) of the subjects in the IAsp group, 41% (24/59) of the subjects in the BR group, and 36% (10/28) of the subjects in the lispro group; between-treatment differences were not statistically significant. There were no episodes of diabetic ketoacidosis during the trial. No clinically significant differences between treatment groups were noted for vital signs, physical parameters, results of electrocardiography, or clinical laboratory findings.
Pump compatibility
The present study indicates that IAsp is as effective as BR and lispro when used in CSII therapy. Most subjects entered into the trial with good glycemic control, as demonstrated by mean baseline HbA1c values ranging from 7.3 to 7.5%, and maintained such control over the 4 months of the trial. These findings confirm the results of an earlier 7-week study comparing IAsp and BR in CSII therapy, in which both treatments provided comparable glycemic control in patients with type 1 diabetes (8). In a CSII study reported by Zinman et al. (7), the HbA1c values for lispro-treated subjects were significantly improved compared with those subjects treated with regular human insulin. However, subjects in that study administered bolus doses of insulin (lispro or regular human insulin) 0 to 5 min before meals; such a study design put regular human insulin at a disadvantage, because it should be given 30 min before mealtime for optimal glucose-lowering effect. Subjects in the Zinman study also had a higher mean baseline HbA1c value (8.0%) than those in the present study (7.37.5%). After 3 months of treatment, the mean HbA1c value decreased to 7.66%, a value that was slightly higher than the mean baseline HbA1c values for subjects in the present study. Subjects were safely switched from BR in the run-in period to insulin analogs during the treatment period without having to make any adjustment in insulin dose (total, bolus, or basal) for maintenance of glycemic control during this CSII trial. As seen in studies with MDI therapy, subjects in this study had higher nighttime BG values with the use of rapid-acting insulin analogs than with BR (11,12). Adjustment of the basal insulin dose may enable patients to achieve improvement in nighttime and between-meal glycemic control, which could lead to decreases in HbA1c. However, the nocturnal basal rate should be increased carefully so that the incidence of nocturnal hypoglycemia is not also increased. The insulin analogs in this study showed an advantage over BR in terms of improved postprandial glycemic control. For eight-point BG profiles, the postprandial BG values of the IAsp group tended to be lower than those of the BR group and were similar to those of the lispro treatment group. These results are consistent with the lowered postprandial BG values of patients using insulin analogs as the mealtime bolus in MDI therapy (1214). The impact of postprandial glycemic control on overall glycemic control was addressed in the recent American Diabetes Association position statement stating that postprandial glucose, as well as fasting plasma glucose, and mean plasma glucose, are highly correlated with HbA1c (15). However, the importance of postprandial hyperglycemia per se as a risk factor for late complications of diabetes has not been clearly established. Nevertheless, several studies in nondiabetic and type 2 diabetic subjects have shown an increased risk of mortality and myocardial infarction associated with isolated postprandial hyperglycemia (1622). The overall occurrence of hypoglycemic episodes with BG values <50 mg/dl was similar for the three treatment groups, although there was a trend toward a lower rate of hypoglycemia in the IAsp group compared with the BR and lispro groups (not significant). A similar finding was demonstrated in two published studies comparing lispro with human regular insulin in CSII, in which the rate of blood glucoseconfirmed hypoglycemia for the insulin analog group was slightly but not significantly lower than the rate for regular insulin group (7,23). A significantly lower rate of hypoglycemic episodes based on symptoms alone was reported for subjects in the IAsp group than for subjects in the BR or lispro groups (Table 2). Although subjects in this study are experienced with CSII and are likely to have good hypoglycemic awareness, the rates of hypoglycemia that are based on symptoms alone are subject to possible bias in an open-label study and could represent symptoms not caused by hypoglycemia. Nocturnal hypoglycemia is particularly problematic in patients with diabetes. In the present study, the IAsp group had a significantly lower rate of nocturnal hypoglycemia compared with the BR group and a similar rate to that of the lispro group (Table 2). A trend of lower rates of nocturnal hypoglycemia with insulin analogs (IAsp and lispro) has also been shown in studies using MDI therapy (24,25). Prolonged clogs and blockages of the pump or infusion sets can result in insufficient administration of insulin, which can potentially lead to life-threatening diabetic ketoacidosis. Crystal formation of soluble insulins can increase the chance of occlusion in pumps and infusion sets. IAsp is a soluble insulin that has been shown to have significantly less crystal formation than BR (8). In the present study, the suitability of IAsp for pump use was similar to that of BR and lispro, as demonstrated by a low incidence of clogs and blockages for all treatment groups. Patients in this trial were experienced with CSII and were properly educated to recognize and correct any clogs and blockages. Accordingly, patients took corrective actions, such that only a small percentage of clogs and blockages (9 to 6%) coincided with hyperglycemia (BG >350 mg/dl); in no case did a hyperglycemic episode progress to diabetic ketoacidosis. The rapid-acting nature of insulin analogs provides patients with greater flexibility of their mealtime insulin needs because the bolus can be administered immediately before meals. This greater flexibility will likely lead to improved compliance and a better quality of life for patients using CSII therapy. In conclusion, IAsp is a safe and effective alternative to BR and lispro for patients with type 1 diabetes and is compatible with pump use in CSII therapy.
Acknowledgments This study was sponsored by Novo Nordisk Pharmaceuticals, Princeton, NJ. The insulin aspart CSII pump study group also includes the following investigators: James Desemone (Albany, NY), Richard Guthrie (Wichita, KS), Alan Marcus (Laguna Hills, CA), Jennifer Marks (Miami, FL), Gregory Peterson (Des Moines, IA), and Jay Skyler (Miami, FL). We thank the following individuals for technical assistance: Deborah Armstrong, Gail Baillargeon, Lauren Currier, Donna Flanders, Kelly McCulloch, Laura Plummer, Carol Racklein, Tammy Stitch, Genene Streimer, Suzanne Strowig, and Marianne Vetrano.
Address correspondence and reprint requests to Bruce W. Bode, MD, Atlanta Diabetes Associates, 77 Collier Rd. NW, Suite 2080, Atlanta, GA 30309. E-mail: bbode001 {at}aol.com. Received for publication 18 June 2001 and accepted in revised form 21 September 2001. B.B. and J.M. has received honoraria from Novo Nordisk Pharmaceuticals. D.N. has received honoraria from Minimed. P.R. has received consultantships and/or is on the speakers board for Asta Medica, Aventis, Bayer, Bristol Myers Squibb, Eli Lilly, Merck, Novo Nordisk, Pfizer, Sanwa, SmithKline Beecham, Takeda America, and Therasense and holds stock options in Therasense. J.A.D. has received consultantships and/or is on the speakers board for Bayer Pharmaceuticals, Bristol Myers Squibb, Eli Lilly, Merck, BioControl Technologies, Pfizer, Novo Nordisk, Dura Pharmaceuticals, SmithKline Beecham, and Takeda. R.H. is a paid consultant for Novo Nordisk. A table elsewhere in this issue shows conventional and Système International (SI) units and conversion factors for many substances.