Clinical Evaluation of Alternative-Site Glucose Measurements in Patients After Major Cardiac Surgery
1 Department of Internal Medicine, Medical University Graz, Graz, Austria Address correspondence and reprint requests to Martin Ellmerer, Department of Internal Medicine, Medical University Graz, Stiftingtalstr. 24, A-8010 Graz, Austria. E-mail: martin.ellmerer{at}healthgate.at
OBJECTIVETight glycemic control improves outcome in critically ill patients but requires frequent glucose measurements. Subcutaneous adipose tissue (SAT) has been characterized as promising for glucose monitoring in diabetes, but it remains unknown whether it can also be used as an alternative site in critically ill patients. The present study was performed to clinically evaluate the relation of glucose in SAT compared with arterial blood in patients after major cardiac surgery. RESEARCH DESIGN AND METHODSForty critically ill patients were investigated at two clinical centers after major cardiac surgery. Arterial blood and SAT microdialysis samples were taken in hourly intervals for a period of up to 48 h. The glucose concentration in dialysate was calibrated using a two-step approach, first using the ionic reference technique to calculate the SAT glucose concentration (SATg) and second using a one-point calibration procedure to obtain a glucose profile comparable to SAT-derived blood glucose (BgSAT). Clinical validation of the data was performed by introducing data analysis based on an insulin titration algorithm. RESULTSCorrelation between dialysate glucose and blood glucose (median 0.80 [interquartile range 0.680.88]) was significantly improved using the ionic reference calibration technique (SATg vs.blood glucose 0.90 [0.830.94]; P < 0.001). Clinical evaluation of the data indicated that 96.1% of glucose readings from SAT would allow acceptable treatment according to a well-established insulin titration protocol. CONCLUSIONSThe results indicate good correlation between SATg and blood glucose in patients after major cardiac surgery. Clinical evaluation of the data suggests that with minor limitations, glucose from SAT can be used to establish tight glycemic control in this patient group.
Abbreviations: BgSAT, subcutaneous adipose tissuederived blood glucose ICU, intensive care unit SAT, subcutaneous adipose tissue SATg, SAT glucose concentration
Maintaining critically ill patients within strict glycemic limits can dramatically reduce mortality, risk of infection, and other complications and also has substantial socioeconomic impact (13). Due to administration of varying doses of parenteral and enteral nutrition, intravenous infusion of medications that affect glucose metabolism, and development of acute insulin resistance during sepsis, tight glycemic control can only be granted by frequent blood glucose monitoring. Although most critically ill and hospitalized patients have routinely placed venous or arterial access lines, glycemic control is still inadequate. The unmanageable workload for the nursing staff and the prevalent fear of hypoglycemia among critical care physicians hinders the implementation of glycemic control in the intensive care unit (ICU) (4). For diabetic patients, alternative-site glucose testing to achieve better, continuous, or automated feedback control of glycemia has been sought for a long time (57). Thanks to minimal invasiveness and good correlation with blood glucose, subcutaneous adipose tissue (SAT) is probably the most investigated alternative sampling site (813). Glucose monitoring in SAT would provide more frequent information about glycemia (trend information) and would therefore also be of benefit for critically ill patients. Microdialysis of SAT has been well established for the investigation of the interstitial fluid glucose profile in relation to blood glucose in healthy individuals and diabetic patients (9). Due to critical illness of patients in the ICU, the perfusion status of various tissue regions is altered (1416), and hence, the suitability of adipose tissue as an alternative sampling site for glucose also has to be questioned. The aim of the present study was to clinically evaluate the relation between glucose concentrations in SAT and arterial blood using hourly point-to-point analysis in postcardiac surgery patients at the ICU for a period of up to 48 h. A novel data analysis was introduced to clinically evaluate whether glucose measurements from SAT can be used to establish tight glycemic control in this patient group.
A total of 40 adult patients from two different ICUs were included in the study. All patients were investigated after major cardiac surgery (coronary artery bypass grafting 70%, valve replacement 15%, both 5%, and aortic replacement 10% of patients). Patients were investigated until the end of ICU stay but for a maximum period of 48 h (median 48 h [range 1548]). Patient characteristics are depicted in Table 1. The study was approved by the local ethics committee at Charles University Prague and at Medical University Graz. Signed informed consent was obtained from all patients before surgery and before any trial-related activities. The study started with a screening visit before surgery to obtain characteristic information about the patients, including demographic data, medical history, concomitant medication, body composition, vital signs, and laboratory analysis from routine laboratory assessment. Patients fulfilling the inclusion criteria (increased blood glucose levels at admission to the ICU [>120 mg/dl]) were included after admission to the ICU. Tight glycemic control was established based on arterial blood glucose measurements. For sampling of interstitial fluid a microdialysis catheter (CMA 60; CMA Microdialysis, Solna, Sweden) was inserted into the SAT on the left or right side of the umbilical region. After insertion, the catheter was connected to a microinfusion pump (CMA 107; CMA Microdialysis) (17,18). Experiments started at least 60 min after insertion of the microdialysis catheter. The dialysis catheter was constantly perfused at a flow rate of 1 µl/min with an isotonic solution of 5% mannitol (19). Dialysate was continuously sampled and collected in hourly fractions (e.g. from time 0800 until 0900) in interstitial sampling vials (CMA Microdialysis) throughout the study period. Blood samples from an arterial line were collected in hourly intervals (e.g. at time 0800 and 0900) at the same time when SAT sampling vials were changed. Experiments ended when patients were transferred from the ICU upon decision of the treating physician but at the latest 48 h after start of the study.
Arterial blood glucose samples were analyzed using standard point-of-care testing devices (Graz: Omni S; Roche Diagnostics, Basel, Switzerland; Prague: ABL 700; Radiometer Medical, Copenhagen, Denmark). Fractionized samples from SAT dialysate were stored at 70° C immediately after sampling and analyzed at a central laboratory (Joanneum Research, Graz, Austria). Dialysate samples were analyzed for glucose using a Cobas Mira Analyzer (Roche Diagnostics, Basel, Switzerland) using standard enzymatic assays (Roche Diagnostics, Mannheim, Germany). Sodium (dialysate sodium) and potassium were analyzed using a flame photometer (Instrumentation Laboratory, Vienna, Austria). Intrarun (interrun) coefficients of variation were 1.3% (1.6%) and 0.4% (0.4%) for glucose and sodium, respectively.
Calculations and calibration
Statistical analysis
Clinical evaluation (insulin titration error grid analysis)
A total of 40 patients were recruited into the trial. In none of the patients was reinsertion of a microdialysis system due to technical failure required. No adverse events related to the interstitial fluid glucose sampling occurred.
Comparison of dialysate and ionic reference calibrated (SATg) glucose profiles
After applying the second calibration step (one-point calibration), the relative differences between arterial blood glucose readings and BgSAT were calculated. Results indicated median 4.8% (interquartile range 14.7 to 2.9) difference between blood glucose and BgSAT readings. Figure 2 depicts eight representative patients, four patients with smallest (Fig. 2A) and four patients with largest (Fig. 2B) relative difference between blood glucose and BgSAT.
Clinical evaluation of BgSAT profiles Clinical evaluation of BgSAT in comparison to blood glucose was performed using insulin titration error grid analysis as described in Fig. 1A. The analysis is based on a total of 1,415 pairs of blood glucose and BgSAT readings from 40 patients. A total of 96.1% of the interstitial fluid glucose readings were found in the acceptable treatment zone; 3.8% of the data were found in the unacceptable violation zone, and 0.1% of the readings were found in the major violation zone (Fig. 1B). Results of the analysis indicated that for a total of 3.9% of the glucose readings from SAT, safe implementation of tight glycemic control using the applied insulin guideline cannot be guaranteed.
With the present study it was demonstrated that in patients recovering from major cardiac surgery at the ICU, arterial blood glucose fluctuations are well described using interstitial fluid glucose measurements from SAT. Clinical evaluation of the data based on a well-accepted insulin titration guideline indicated that 96% of the glucose readings as derived from SAT would allow establishing tight glycemic control in the investigated patient group. Numerous studies have been performed to test the suitability of glucose monitoring in interstitial fluid of SAT aimed to establish improved glycemic control in diabetes therapy (12,2426). To date, few studies are available that evaluate the subcutaneous glucose concentration in critically ill patients (3,18), and to the best of our knowledge no study with point-to-point analysis of blood versus interstitial fluid glucose readings has been performed so far. Studies investigating more methodological aspects of tissue perfusion in critically ill patients with sepsis indicated impaired microvascular perfusion in these patient groups (1416), which also might have an effect on the relation between blood and interstitial glucose concentrations. Results from the present study in patients at the surgical ICU suggest a close relationship between blood and interstitial fluid glucose, which is comparable as seen in previous studies in healthy volunteers and type 1 diabetic patients (9,12). The present study is part of CLINICIP (Closed Loop Insulin Infusion for Critically Ill Patients; www.clinicip.org), an integrated project funded by the European Community. As part of the project, the present study was performed to directly evaluate the principle relation between arterial blood and interstitial fluid glucose concentrations using state-of-the-art laboratory analysis and point-to-point comparisons. Basis for the data analysis of the present study were samples of interstitial fluid dialysate, which were continuously collected in hourly fractions. A prospective calibration technique (ionic reference technique) was used to calculate the interstitial fluid glucose concentration (19). To avoid the necessity of individual plasma sodium measurements, plasma sodium was assumed constant, a prerequisite that is likely to be violated in critically ill patients. In the present study a maximum intrapatient coefficient of variation (CV) for plasma sodium of 2.3% and an interpatient CV of 4.6% was observed. Regardless of these fluctuations of plasma sodium, the correlation between interstitial fluid and blood glucose readings could be substantially improved using the ionic reference technique. This finding clearly suggests that the calibration technique is useful when applied in the investigated patient group. For the clinical evaluation of glucose data from SAT, a novel approach, an insulin titration error grid analysis was introduced. The objective of the analysis is to provide clinical evaluation criteria for alternative-site versus reference glucose measurements for critically ill patients. The analysis is based on a widely used insulin titration protocol as proposed by Van den Berghe (22). To integrate the insulin titration protocol into a standard two-dimension x-y graph, the titration guideline was simplified with respect to actions per range and at the same time also slightly extended, i.e., an additional action for glucose exceeding 300 mg/dl was introduced. Analysis of the present data using the analysis indicated that a total of 3.9%, representing 55 of 1,415 blood glucose readings, were outside the clinical acceptable treatment zone. The data suggest that establishment of tight glycemic control according to the implemented guideline and based on glucose readings from interstitial fluid would not have been violated for 96% of the acquired data pairs. Further analysis indicated that data in the unacceptable treatment zone was mostly due to patients, as indicated in Fig. 2B. Instability of the microdialysis catheter (25) during the initial phase of the experiment in combination with the applied one-point calibration procedure might explain differences of the subcutaneous glucose signal compared with arterial blood glucose readings as seen in these patients. Additional refinement of the technology (e.g., more frequent calibration intervals) will be required to further improve the relation between blood and interstitial glucose readings. We summarize that the relation between arterial blood glucose and glucose as derived from interstitial fluid measurements was similar in patients after major cardiac surgery as previously reported for healthy volunteers and type 1 diabetic patients. Critical evaluation of the interstitial glucose data suggests that with minor limitations, glucose from SAT can be used to establish tight glycemic control in patients after major cardiac surgery.
The study is part of CLINICIP (Closed Loop Insulin Infusion for Critically Ill Patients), an IST (Information Society and Technology) project funded by the European Community under the Sixth Framework Program, Action Line eHealth, Project Reference 506965. The authors thank Agnes Mautner and Maria Suppan from JOANNEUM Research, Graz, Austria, for expert analysis of microdialysis samples.
A table elsewhere in this issue shows conventional and Système International (SI) units and conversion factors for many substances. The costs of publication of this article were defrayed in part by the payment of page charges. This article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact. Received for publication December 5, 2005. Accepted for publication February 16, 2006.