Subcutaneous Glucose Concentration in Humans: Real estimation and continuous monitoring
- Fabio Sternberg, MD,
- Carsten Meyerhoff, DR MED,
- Franz J Mennel, DR RER NAT,
- Friederike Bischof, DR MED and
- Ernst F Pfeiffer, DR MED
- Address correspondence and reprint requests to Fabio Sternberg, MD, Institut fur Diabetes-Technologie an der Universität Ulm, Helmholtzstr. 20, D-89081 Ulm/Donau, Germany.
Abstract
OBJECTIVE To determine the real subcutaneous glucose concentration in healthy volunteers to help in the development of new calibration methods for subcutaneous glucosensors.
RESEARCH DESIGN AND METHODS We developed a new method to estimate the real subcutaneous glucose concentration based on the recirculation of phosphate-buffered saline (PBS) in a microdialysis probe inserted into the subcutaneous tissue. Tissue glucose diffuses into the probe until complete equilibration between the glucose concentration outside and inside the microdialysis probe is achieved. Later, the glucose content of the recirculated PBS is assessed in vitro. We applied the method in 10 healthy volunteers under fasting state and during a hyperglycemic clamp. In addition, we monitored the subcutaneous glucose with an enzymatic-amperometric glucosensor combined with a microdialysis probe.
RESULTS The subcutaneous glucose concentration measured by the recirculation method was 72 ± 6 and 78 ± 6% of the blood glucose measured in the fasting state and during the hyperglycemic clamps, respectively. On the other hand, the glucosensor's signal correlated significantly with the blood glucose.
CONCLUSION The recirculation method estimated the real subcutaneous glucose concentration, opening the way to develop new calibration procedures for subcutaneous glucosensors. However, a suitable calibration procedure is still lacking.
- Received May 4, 1994.
- Revision received May 25, 1995.
- Accepted May 25, 1995.
- Copyright © 1995 by the American Diabetes Association











