Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors
Response to Brauker and Matsubara
- Boris Kovatchev, PHD1,
- Stacey Anderson, MD1,
- Lutz Heinemann, PHD2 and
- William Clarke, MD1
- 1University of Virginia, Charlottesville, Virginia
- 2Profil Institute for Metabolic Research, Neuss, Germany
- Corresponding author: Boris Kovatchev, boris{at}virginia.edu
In their comments on our assessment of the accuracy of four continuous glucose monitoring (CGM) sensors, Brauker and Matsubara (1) correctly point out that our presented studies employed hyperinsulinemic-euglycemic and hypoglycemic clamps, which tested the four devices (Guardian, DexCom 3-day sensor, Navigator, and Glucoday) close to the limits of their stated glucose range. We agree that such a test goes beyond evaluation of the devices per their …