Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors

Response to Brauker and Matsubara

  1. Boris Kovatchev, PHD1,
  2. Stacey Anderson, MD1,
  3. Lutz Heinemann, PHD2 and
  4. William Clarke, MD1
  1. 1University of Virginia, Charlottesville, Virginia
  2. 2Profil Institute for Metabolic Research, Neuss, Germany
  1. Corresponding author: Boris Kovatchev, boris{at}virginia.edu

In their comments on our assessment of the accuracy of four continuous glucose monitoring (CGM) sensors, Brauker and Matsubara (1) correctly point out that our presented studies employed hyperinsulinemic-euglycemic and hypoglycemic clamps, which tested the four devices (Guardian, DexCom 3-day sensor, Navigator, and Glucoday) close to the limits of their stated glucose range. We agree that such a test goes beyond evaluation of the devices per their …