Relationship of A1C to Glucose Concentrations in Children With Type 1 Diabetes

Assessments by high-frequency glucose determinations by sensors

  1. Diabetes Research in Children Network (DirecNet) Study Group*
  1. Address correspondence and reprint requests to Darrell M. Wilson, MD, c/o DirecNet Coordinating Center, Jaeb Center for Health Research, 15310 Amberly Dr., Suite 350, Tampa, FL 33647. E-mail: direcnet{at}jaeb.org

Abstract

OBJECTIVE—Despite the standing of A1C as the most validated and widely used measure for average glycemic control over time, the relationship between A1C and glucose concentrations is not completely understood. The purpose of this Diabetes Research in Children Network (DirecNet) study was to use continuous glucose monitoring data to examine the relationship between A1C and glucose in type 1 diabetes.

RESEARCH DESIGN AND METHODS—Forty-eight youth enrolled in studies of the Navigator continuous glucose monitor were encouraged to wear the Navigator sensor at home continuously. A1C was measured at baseline, at 3 months, and at 6 months. Sensor glucose data were directly transmitted via the Internet, assuring that essentially all glucose values were analyzed.

RESULTS—Subjects had a median of 112 h/week of Navigator data in the first 3 months and 115 h/week in the second 3 months. The slope of mean glucose over the previous 3 months versus A1C was only 18 mg/dl per 1.0% A1C. Individually, there was substantial variation in the relationship between mean glucose and A1C. A1C was not associated with glucose lability after controlling for mean glucose. Measures of an individual's rate of glycation were moderately correlated at the 3- and 6-month visits.

CONCLUSIONS—As the chemistry of glycation would predict, we found no evidence to contradict the simple hypothesis that A1C directly reflects mean glucose over time. There is, however, substantial variability in individual mean glucose concentrations for a given A1C. Transforming reliable A1C values into calculated mean glucose values would, when applied to an individual, introduce substantial error.

Despite the standing of A1C as the most validated and widely used measure for average glycemic control over time, the relationship between A1C and blood glucose concentrations is not completely understood. The chemistry of this first order, nonenzymatic glycation reaction should reflect the integral of glucose concentration over time, adjusted for the survival of erythrocytes, but this simple relationship has been questioned by many clinicians and investigators, as discussed by Cohen et al. (1). Some clinicians hold that higher glucose concentrations increase the A1C more than lower glucose levels reduce it. Some studies have indicated that postprandial glucose values have a greater impact on A1C than do preprandial glucose concentrations (2), whereas others suggest that the opposite is true (3). Even if we assume that A1C linearly reflects a time weighted average of blood glucose concentrations, the exact ratio between average blood glucose and A1C remains uncertain, with published data indicating that a 1 percentage point rise in A1C is the equivalent to anywhere from a 19 to 36 mg/dl rise in average glucose concentrations (4,5).

Many factors contribute to different estimates of the slope of the relationship between average plasma glucose and A1C. One major reason for this uncertainty stems from limitations in determining the average glucose level using traditional conventional home glucose monitoring systems. Typically, glucose is measured infrequently, and patients often oversample when they are symptomatic or when a measurement is outside their target range and undersample at night.

Another unresolved and controversial issue is whether individuals glycate hemoglobin proteins at different rates. Among 223 adults without diabetes, differences in glucose intolerance explained only one-third of the variance found in glycated hemoglobin levels (6). Comparing nondiabetic monozygotic and dizygotic twins, Cohen et al. (7) estimated that 69% of the variance in the glycation gap was heritable. This question has taken on added importance because of the recent suggestion that higher glycation rates may predict increased risk for long-term complications independently of glycemic control (8), although this suggestion is controversial (9).

The Diabetes Research in Children Network (DirecNet) has carried out two nonrandomized studies to evaluate the efficacy of the FreeStyle Navigator Continuous Glucose Monitoring System (Abbott Diabetes Care, Alameda, CA) in children and adolescents with type 1 diabetes treated with insulin pump and glargine-based multiple daily injection (MDI) regimens. These studies involved subjects whose A1C levels varied widely. The wealth of continuous glucose monitoring data that was generated by these 6-month studies provided us with a unique opportunity to examine the relationship between A1C and glucose concentrations in youth with diabetes.

RESEARCH DESIGN AND METHODS

The procedures for the Navigator studies have been described in detail elsewhere (10) and are briefly summarized here. Fifty-seven subjects with type 1 diabetes, aged 4-<18 years with A1C 5.8–10.3%, were enrolled into nonrandomized pilot studies of Navigator use in pump (n = 30) and glargine-based MDI (n = 27) regimens; 93% of the subjects were non-Hispanic whites. This system combines a glucose sensor with a built in FreeStyle glucose meter, which was used to calibrate the Navigator four times, at ∼10, 12, 24, and 72 h after insertion. Subjects were encouraged to wear the Navigator sensor at home as often as possible. A1C was measured at baseline, at 3 months, and at 6 months with the DCA 2000+ Analyzer (Bayer, Tarrytown, NY). We have previously shown that A1C levels measured by the DCA2000+ are highly correlated (r = 0.94) with reference A1C levels measured in the Diabetes Control and Complications Trial (DCCT)/Epidemiology of Diabetes Interventions and Complications (EDIC) Laboratory at the University of Minnesota (Minneapolis, MN), albeit with slightly higher values (mean difference +0.2% vs. DCCT/EDIC Laboratory with 95% CI of +0.14 to 0.23%, P < 0.001) (11). The median relative absolute difference was 0.03. Sensor glucose data were directly transmitted from the subject's home to the DirecNet Coordinating Center, assuring that essentially all glucose values were available for analysis.

Statistical methods

For analysis of mean glucose measured by the Navigator sensor versus A1C measured by the DCA2000+ data from the 3- and 6-month visits were combined. A visit was included in the analysis if the subject had at least 24 h of Navigator data for at least 6 (not necessarily consecutive) of the 13 prior weeks. One outlier at the 3-month visit with a mean glucose of 233 mg/dl and an A1C of 8.3% was excluded from analysis (results were similar whether or not this data point was included). Forty-seven subjects (28 pump users and 19 MDI-glargine users) met this criterion for the 3-month visit and 28 subjects (17 pump users and 11 MDI-glargine users) met this criterion for the 6-month visit. The combined analysis included 75 visits from 48 subjects. All statistical analyses were conducted using SAS version 9.1 software (SAS Institute, Cary, NC).

The mean glucose values from the sensor data were calculated, giving equal weight to each of the 24 h of the day. Repeated-measures least-squares regression was performed using mean glucose as the dependent variable and A1C as the independent variable, combining data from the 3- and 6-month visits. There were no significant interactions with visit time (3 vs. 6 months) or insulin modality (pump vs. MDI).

Two measures of glycation were calculated from the sensor data: first a simple ratio of mean glucose to A1C and second the hemoglobin glycation index defined as the residual from a regression model with A1C as the dependent variable and mean glucose as the independent variable (8). These measures are intended to quantify an individual's tendency to glycate and may be helpful in predicting an individual's future risk for complications. Pearson correlation was used to assess the consistency of these measures between 3 and 6 months.

Additional regression models were run with A1C as the dependent variable to see whether any measure of glucose lability was predictive after adjustment for mean glucose. Lability measures included mean amplitude of glycemic excursions (12), absolute rate of change (13), and percentage of glucose values >250 and >300 mg/dl. Separate models were run to test these lability measures one at a time, adjusting for mean glucose.

RESULTS

The 47 subjects (28 pump users and 19 MDI-glargine users) whose 3-month visits were included in the analysis had a median (25th–75th percentiles) 112 (83–130) hours of Navigator data per week. The 28 subjects (17 pump users and 11 MDI-glargine users) whose 6-month visits were included in the analysis had 115 (88–131) hours of Navigator data per week.

The slope ± 95% CI of mean glucose over the previous 3 months versus A1C was 18 ± 4 mg/dl per 1.0% A1C. Moreover, for a given A1C, mean glucose concentrations varied considerably (Fig. 1). For example, for an A1C of 7%, mean glucose concentrations ranged from 138 to 189 mg/dl. This relation was similar for subjects using a pump or MDI-glargine and at the 3- and 6-month visits (Fig. 1).

Measures of an individual's rate of glycation were moderately correlated at the 3- and 6-month visits (Fig. 2). The Pearson correlation was 0.70 (95% CI 0.42–0.85) for the ratio of mean glucose (over the previous 3 months) to A1C and 0.63 (0.33–0.81) for the hemoglobin glycation index. It is noteworthy that the slope of the relationship between the change in mean sensor glucose and change in A1C between 3 and 6 months was 12 ± 9 mg/dl (Fig. 3).

There was no evidence to suggest that the rate of glycation changed nonlinearly with glucose concentration. After we controlled for the mean glucose over the prior 3 months, A1C was not significantly associated with measures of glucose lability (mean amplitude of glycemic excursions P = 0.33, absolute rate of change P = 0.76, percentage of glucose sensor readings >250 mg/dl P = 0.18, and percentage of glucose sensor readings >300 mg/dl P = 0.39).

CONCLUSIONS

By the design of our data capture system, we were able to collect and analyze essentially all sensor glucose values obtained by our subjects. Just as chemistry of glycation would predict, we found no evidence to contradict the simple hypothesis that A1C directly reflects the integral of glucose level over time, often referred to as the area under the curve. Thus, high glucose values, such as those typically seen postprandially, do not disproportionally contribute to A1C. Moreover, glucose variability does not appear to affect the A1C. These results are similar to those of Salardi et al. (14) using the retrospective continuous glucose monitoring system device from the Medtronic MiniMed.

The slope of the regression equation relating A1C to mean glucose in this study (18 mg/dl for each 1 percentage change in A1C) is lower than that found by others. Rohlfing et al. (4), using data from the DCCT, found a slope of 36 mg/dl. In a recent small study, Nathan et al. (15) found a tighter association between mean glucose and A1C and a slope of 32 mg/dl. Their adult subjects included patients with both type 1 and type 2 diabetes as well as subjects without diabetes. The slope observed in this study is consistent with that we reported previously in 200 youths with type 1 diabetes involved in the DirecNet randomized clinical trial of the GlucoWatch G2 Biographer, which was based on either 3 days of 8-point meter testing (i.e., 1% A1C change for every 23 mg/dl change in mean plasma glucose) or 3 days of continuous glucose monitoring using the MiniMed continuous glucose monitoring system (i.e., 1% A1C change for every 19 mg/dl change in mean plasma glucose) (5).

Of note, the subjects in the DirecNet studies are children and adolescents who may have higher glycemic variability than adults. Our findings may not therefore be directly applicable to adult patients. The difference in slope, however, has implications for the care of patients with diabetes. A smaller improvement in average glucose concentration translates, on average, to a larger improvement in A1C. Moreover, in this study there is substantial variability in individual mean glucose concentrations for a given A1C. For any given A1C level, mean sensor glucose levels differed by up to 50 mg/dl or more (Fig. 1), making the conversion of A1C levels into mean glucose equivalents as suggested by a recent American Diabetes Association consensus statement (16) tenuous at best.

An additional aim of this study was to use our continuous glucose monitoring data to explore whether and to what extent rates of glycation or turnover of glycated hemoglobin vary among individual patients. It is also particularly important that both measures of glycability were reasonably constant over the 6 months of our study, consistent with the hypothesis that individual subjects with diabetes glycate proteins at differing rates and that this tendency to be a fast or slow glycator persists over time within an individual. However, individual differences in glycability explained <50% of the differences in A1C levels among patients, and a number of other factors, such as differences in erythrocyte turnover or the interassay coefficient of variation of A1C measurements, undoubtedly play a role. Herman et al. (17), using data from the Diabetes Prevention Program, have recently reported that A1C, adjusted for glycemic indexes, varies by race and ethnicity.

Within our DirecNet trials, we have demonstrated that A1C measurement, a recognized valid measure of diabetes control, can be quite reliable. We found that A1C measurements obtained with the Bayer DCA2000+ matched those for the reference method closely (11). Some have suggested that transforming these reliable A1C values into calculated mean glucose values would improve diabetes management. Although A1C clearly reflects mean glucose, our data demonstrate that this approach, when applied to an individual, could introduce substantial error. Moreover, A1C may be such a strong surrogate marker for the risk of diabetic complications because it reflects both the average plasma glucose level and the propensity of the patient to glycosylate other structural proteins. More long-term studies are needed to assess the clinical meaning of these findings.

APPENDIX

Writing Committee

Darrell M. Wilson, MD; Craig Kollman, PhD; Dongyuan Xing, MPH; Rosanna Fiallo-Scharer, MD; Stuart Weinzimer, MD; Michael Steffes, MD, PhD; Tim Wysocki, PhD; Roy Beck, MD, PhD; Katrina Ruedy, MSPH; William Tamborlane, MD; and the Diabetes Research in Children Network (DirecNet) Study Group.

The DirecNet Study Group

Clinical centers are listed in alphabetical order with clinical center name, city, and state. Personnel are listed as “PI” for principal investigator, “I” for co-investigator and “C” for coordinators. 1) Barbara Davis Center for Childhood Diabetes, University of Colorado, Denver, CO: H. Peter Chase, MD (PI), Rosanna Fiallo-Scharer, MD (I), Laurel Messer, RN (C), and Barbara Tallant, RN, MA (C); 2) Department of Pediatrics, University of Iowa Carver College of Medicine, Iowa City, IA: Eva Tsalikian, MD (PI), Michael J. Tansey, MD (I), Linda F. Larson, RN (C), Julie Coffey, MSN (C), and Joanne Cabbage (C); 3) Nemours Children's Clinic, Jacksonville, FL: Tim Wysocki, PhD, ABPP (PI), Nelly Mauras, MD (I), Larry A. Fox, MD (I), Keisha Bird, MSN (C), and Kim Englert, RN (C); 4) Division of Pediatric Endocrinology and Diabetes, Stanford University, Stanford, CA: Bruce A. Buckingham, MD (PI), Darrell M. Wilson, MD (I), Jennifer M. Block, RN, CDE (C), Paula Clinton, RD, CDE (C), and Kimberly Caswell, APRN; 5) Department of Pediatrics, Yale University School of Medicine, New Haven, CT: Stuart A. Weinzimer, MD (PI), William V. Tamborlane, MD (I), Elizabeth A. Doyle, MSN (C), Heather Mokotoff, MSN (C), and Amy Steffen (C); Coordinating Center: Jaeb Center for Health Research, Tampa, FL: Roy W. Beck, MD, PhD, Katrina J. Ruedy, MSPH, Craig Kollman, PhD, Dongyuan Xing, MPH, Mariya Dontchev, MPH, Cynthia R. Stockdale, and Judy Jackson; University of Minnesota Central Laboratory: Michael W. Steffes, MD, PhD, Jean M. Bucksa, CLS, Maren L. Nowicki, CLS, Carol A. Van Hale, CLS, and Vicky Makky, CLS; National Institutes of Health: Gilman D. Grave, MD, Mary Horlick, PhD, Karen Teff, PhD, and Karen K. Winer, MD; Data and Safety Monitoring Board: Dorothy M. Becker, MBBCh, Patricia Cleary, MS, Christopher M. Ryan, PhD, Neil H. White, MD, CDE, and Perrin C. White, MD.

Figure 1—

Mean glucose versus A1C. Mean glucose measured by the Navigator sensor over a 3-month period before the A1C measurement (n = 75 visits from 48 subjects). The regression line was calculated from a repeated-measures model accounting for correlated data from the same subject.

Figure 2—

Measures of glycation at 3- versus 6-month visits. n = 27 subjects with data at both visits. The ratio of mean glucose to A1C (A) and the hemoglobin glycation index (B) are compared at the two visits. The diagonal represents the line of identity.

Figure 3—

Change in mean glucose versus change in A1C from 3 to 6 months with regression line.

Acknowledgments

This research was supported by the National Institutes of Health/National Institute of Child Health and Human Development Grants HD041919–01, HD041915–01, HD041890, HD041918–01, HD041908–01, and HD041906–01. Clinical centers also received funding through General Clinical Research Center Grants Numbers M01 RR00069, RR00059, RR 06022, and RR00070-41. Abbott Diabetes Care, Alameda, CA, provided the FreeStyle Navigator Continuous Glucose Monitoring Systems and the FreeStyle Blood Glucose Meter test strips.

Appreciation is expressed for the work performed by the clinical research center nurses at the five clinical centers.

Footnotes

  • Published ahead of print at http://care.diabetesjournals.org on 4 December 2007. DOI: 10.2337/dc07-1835.

  • *

    * A complete list of the members of the DirecNet Study group can be found in the appendix.

  • The costs of publication of this article were defrayed in part by the payment of page charges. This article must therefore be hereby marked “advertisement” in accordance with 18 U.S.C Section 1734 solely to indicate this fact.

    • Accepted November 23, 2007.
    • Received September 18, 2007.

References

| Table of Contents

This Article

  1. Diabetes Care vol. 31 no. 3 381-385
  1. All Versions of this Article:
    1. dc07-1835v1
    2. 31/3/381 most recent