Pharmacokinetics and Pharmacodynamics of Biosimilar Insulins: Is Clamp Technology Fit for Purpose?
- Philip Home⇑
- Corresponding author: Philip Home, philip.home{at}ncl.ac.uk.
The euglycemic glucose clamp was first used to assess the new human insulins in the early 1980s by comparing pharmacodynamics (PD) of square-wave intravenous infusions in terms of the clamp glucose infusion rate (1). Shortly afterward, the technique was adapted to study the profile of action of subcutaneously injected insulin (2) and in people with type 1 diabetes rather than people without diabetes. One major advantage of the clamp is that it allows the study at the same time of both the pharmacokinetics (PK), the plasma concentration absorption profile of injected insulin, and PD in terms of measures such as onset, peak, peak:trough ratio, duration, and area under the curve (Table 1). Attractively, the time-action profiles can be visualized. Although initially used for mealtime insulins, the technique was soon applied to extended-acting insulins, comparing insulin-action profiles of nominally similar insulins (NPH preparations) from different manufacturers (3).
Some glucose clamp advantages and limitations in the study of subcutaneous injection profiles of insulin preparations
Mealtime insulins can be compared, to a useful extent, in people with type 1 diabetes (C-peptide negative) using standardized or usual meals on a background of steady-state basal insulin or infusion (4). Such an option is not available for long-acting insulins, so glucose clamp studies are particularly important for assessing and comparing these. Thus, while general guidelines for comparing a biosimilar medication (a copy) with the original product emphasize PK/PD studies (5,6), specific guidelines for insulin as drawn up by the European regulators go into considerable detail on glucose clamp studies (7). In this issue of Diabetes Care, Linnebjerg et al. (8), in an article comparing the Lilly glargine preparation to the reference Sanofi glargine, report three glucose clamp studies in people without diabetes. Additionally, studies have been performed in people with type 1 diabetes and to assess dose ranging (9,10).
Although we have three decades of experience in the techniques used, debate still rages on methodology and interpretation of the data—indeed for both the PK and PD findings. Some of the issues are discussed in the article by Swinnen et al. (11), written after experts in the field failed to reach consensus over methodology, and some advantages and issues are listed in Table 1.
The PK issues are perhaps the more surprising. What could be easier than measuring insulin concentrations after injection for 24 h or more while fasting, indeed even in the absence of a clamp? But first, if people without diabetes are studied (or more problematically those with type 2 diabetes), then allowance has to be made for changing endogenous insulin secretion. Insulin:C-peptide ratios are usually used for this purpose, referenced to the preinjection baseline ratio. Although C-peptide has a much longer plasma half-life (20 min) than insulin (4 min), this approach will generally remain valid, provided the clamp is of good quality (absence of rapid up-and-down glucose concentration excursions changing the stimulus to endogenous secretion) (Table 1) (12). However in the study by Linnebjerg et al. (8), while the mean PK profiles are identical between the insulins despite high variance at all time points, the recorded profile is not credible, jumping from zero (before injection) to half maximum in just 30 min, very different from what is expected for insulin glargine (13). This unexplained bias probably continues for the full 24 h of study, as the levels at 24 h are higher than might be expected from our knowledge of the insulin effect in clinical practice. Some of the problem here may be that many of the measurements made are below the lower level of quantification (LLoQ) of the insulin assay, and indeed the mean level at many time points is just twice that level. The authors imputed a level of half the LLoQ for any result below it and then discarded any result where the endogenous component calculated from C-peptide resulted in negative findings for insulin glargine. These actions would certainly bias the findings throughout the profile in unpredictable ways, mostly at the start and end of the clamp, the periods of most interest.
This might seem to point to an advantage for clamps in people with type 1 diabetes. Unfortunately, this is not the case. Initially and for some hours thereafter people with no endogenous insulin secretion need a tapering insulin infusion to maintain glucose control while absorption of the long-acting insulin begins (Table 1) (14), and this again becomes true late in the clamp in at least some individuals (although some researchers allow hyperglycemia to develop at this point). Furthermore, even low levels of insulin antibodies can interfere with the insulin assays, and perhaps it is this problem that led to the very erratic findings in the Lilly type 1 diabetes biosimilar clamp studies of very wide CIs on the PK data between insulins essentially making the assay data noncontributory to the assessment of biosimilarity (9,10).
The problem of the need for exogenous insulin at the start and end of the clamps in people with type 1 diabetes also bugs the glucose infusion (PD) data (Table 1). As noted above, some researchers have not given extra insulin at those times, but the developing hyperglycemia then means the remaining glucose-lowering effect of the studied insulin is not quantified, being counted as zero or even censored (14). Furthermore, the lack of endogenous insulin secretion means glucose lowering throughout the clamp is more variable, as it is in type 1 diabetes in clinical practice, again adding to statistical uncertainty and resulting in 95% CIs of ‒54 to 30% for ratio of total glucose infused during the clamp in the Lilly studies (9,10). Furthermore, insulin action in the critical period for extended-acting insulins (18–24 h) will be abnormal in people essentially fasted for that time, meaning the absolute clamp glucose infusion rates will not be a reliable reflection of the glucose-lowering efficacy of the studied insulin at those times (Table 1).
So do the clamps in normal people give confidence that Lilly and Sanofi insulin glargine are clinically similar? For the metrics that probably matter to people with diabetes and clinicians (namely, comparative efficacy at 24 h from injection and the ratio of peak:24-h PK and PD), no data are given, but the published profiles are visually reassuring. The summary data that are given, for example, 24-h area under the curve, are not very helpful—it could be similar for insulins with very different time-action profiles. But even those data do suggest the clamps can only perform to 95% confidence limits of about ±10%, and sometimes much worse (8,10). Clinicians will need to ask themselves whether they are happy that nominally similar (“biosimilar”) insulins could be different by 10%, particularly if substitution within the individual person with diabetes is contemplated or mandated.
Fortunately for the Lilly insulin glargine, other information is available, including clinical studies that ascertained other metrics of clinical importance, such as fasting plasma glucose and hypoglycemia incidence. It is that complete package of data that lends credence to Lilly's claim of biosimilarity rather than the clamp data in isolation (10,15).
Article Information
Acknowledgments. The author is grateful to many colleagues involved in insulin development for lively discussions of some of the issues addressed in this article.
Duality of Interest. P.H., or institutions with which he is associated, receives funding for his research, advisory, and lecturing activities from actual and potential insulin manufacturers of original products or copies, including AntriaBio, Biocon-Mylan, Eli Lilly, Hanmi, Merck, Novo Nordisk, and Sanofi. No other potential conflicts of interest relevant to this article were reported.