Comparison of the lowest and highest nighttime CGMS glucose levels to simultaneous analyzer glucose levels (mg/dl), and the percent difference between them
Patient | Lowest CGMS | Matched analyzer | % Difference | Highest CGMS | Matched analyzer | % Difference |
---|---|---|---|---|---|---|
1 | <40 | 125 | >68 | 131 | 138 | 5 |
2 | 64 | 111 | 42 | 178 | 135 | −32 |
3 | 86 | 109 | 21 | 236 | 253 | 7 |
4 | 45 | 76 | 41 | 185 | 193 | 4 |
5 | 114 | 169 | 33 | 337 | 280 | −20 |
6 | 70 | 103 | 32 | 187 | 167 | −12 |
7 | 47 | 65 | 28 | 195 | 160 | −22 |
Average | 67 ± 26 | 108 ± 34 | 38 ± 15 | 207 ± 65 | 189 ± 57 | −10 ± 15 |
Analyzer glucose levels were considered the standard.